FDA-Device2019-10-30Class II

Torrent Suite Dx Software Version 5.8 Catalog Number:A36601 and A36602 OEM Customer software: Torrent Suite Dx T430 Software v5.0 DARUI DR-SEQ 800 Software UDI:(01)10190302014090 (11)000000(10)5.8 (240)A36601

Life Technologies Holdings Pte Ltd
Hazard

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Sold states
states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland
Affected count
total of Ion+vela +chassis =400 units