FDA-Device2021-11-10Class II
Arietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3
FujiFilm Healthcare Americas Corporation
Hazard
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
Sold states
Nationwide
Affected count
15 units
Manufactured in
1959 Summit Commerce Park, N/A, Twinsburg, OH, United States
Products
Arietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3