FDA-Device2014-11-19Class II

The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Steris Corporation
Hazard

The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.

Sold states
US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.
Affected count
999
Manufactured in
5960 Heisley Rd, Mentor, OH, United States
Products
The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.