FDA-Device2018-10-31Class II

Nuclear Magnetic Resonance Imaging Systems that are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body.

Canon Medical System, USA, INC.
Hazard

When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.

Sold states
Domestic: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY. International: Algeria, Australia, Bangladesh, Brazil, Canada, China, Colombia, Congo, Dominican Republic, Egypt, Guinea, Honduras, India, Indonesia, Iran, Japan, Kenya, Korea, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Nigeria, Oman, Pakistan, Paraguay, Qatar, Saudi Arabia, Singapore, South Africa, Sudan, Taiwan, Thailand, Turkey, UAE, and Vietnam
Affected count
252
Manufactured in
2441 Michelle Dr, N/A, Tustin, CA, United States
Products
Nuclear Magnetic Resonance Imaging Systems that are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0199-2019