FDA-Device2019-11-06Class I
ECHELON FLEX ENDOPATH 60mm Stapler - 340mm shaft, Product Code EC60A
Ethicon Endo-Surgery Inc
Hazard
The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.
Sold states
Worldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.
Affected count
849 units
Manufactured in
4545 Creek Rd, Blue Ash, OH, United States
Products
ECHELON FLEX ENDOPATH 60mm Stapler - 340mm shaft, Product Code EC60A
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0200-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.