FDA-Device2018-10-31Class II

Boston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,

Boston Scientific Corporation
Hazard

Reports of unintended continuation of radiofrequency (RF) energy delivery from the Cardiac ablation system after the foot switch has been released.

Sold states
U.S.: PA, IL, NY, CA, SC, AR, SD, AZ, KY, FL, TX, MA, MI, OH, LA, VA, NC, WA, DE, IN, TN, NJ, CT, AL, ME, GA, NH, MN, MD, DC, OK, HI, WI, CO, MO, IA, GU, NE, OR, ID, UT, RI, KS, NV, MS, WV Foreign (OUS): Canada, Albania, Austria, Belgium, Bulgaria, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Lithuania, Namibia, Netherlands, Lebanon, Pakistan, Poland, Portugal, Russian Fed., Saudi Arabia, South Africa, Spain, Sudan, Sweden, Switzerland, Turkey, Unit.Arab Emir., Argentina, Australia, Bangladesh, Chile, China, Colombia, Japan, Malaysia, Mexico, Nicaragua, Panama, Philippines, Sri Lanka, Taiwan, Thailand, Vietnam
Affected count
793
Manufactured in
100 Boston Scientific Way, N/A, Marlborough, MA, United States
Products
Boston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0202-2019