FDA-Device2023-11-08Class II
Nova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 63368 Stat Profile Prime Plus POC Analyzer (2) PN 63369 Stat Profile Prime Plus POC Analyzer, Remanufactured P/N: 57400, 63368
Nova Biomedical Corporation
Hazard
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
Sold states
Nationwide Foreign: Country Algeria Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Croatia CYPRUS Czech Republic Ecuador Egypt El Salvador Finland France Germany Greece Guatemala India Indonesia Ireland Israel Italy Japan JORDAN Kazakhstan Kenya Kurdistan, Iraq Kuwait Malaysia Malta Mexico Moldova Morocco Nepal Netherlands New Zealand Oman Pakistan Palestine Paraguay Paraquay Peru Philippines Poland Puerto Rico Qatar Republic of Panama Romania Russia Rwanda Saudi Arabia Serbia Slovenia South Africa South Korea Spain Sri Lanka Sudan Sweden Switzerland Syria Thailand Trinidad Trinidad,W.I. Tunisia UK Ukraine United Arab Emerites Uruguay Vietnam
Affected count
161 units
Manufactured in
200 Prospect St, N/A, Waltham, MA, United States
Products
Nova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 63368 Stat Profile Prime Plus POC Analyzer (2) PN 63369 Stat Profile Prime Plus POC Analyzer, Remanufactured P/N: 57400, 63368
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0208-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief