FDA-Device2021-11-17Class II
Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20)
Hazard
When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
Sold states
Domestic distribution nationwide. International distribution worldwide.
Affected count
428 (US); 581 (OUS)
Manufactured in
222 Jacobs St, N/A, Cambridge, MA, United States
Products
Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0213-2022More Philips North America Llc recalls
- FDA-Device2026-07-01Philips Avalon FM20 Fetal Monitor recalled for speaker connector assembly defect
- FDA-Device2026-07-01Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
- FDA-Device2026-07-01Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
- FDA-Device2026-06-17Philips DigitalDiagnost C90 Flex/Value/Chest/ER recalled for missing seismic stabilization kit