FDA-Device2025-10-29Class II

LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.

Synthes (USA) Products LLC
Hazard

Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

Sold states
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
Affected count
122 units
Manufactured in
1301 Goshen Pkwy, West Chester, PA, United States
Products
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0229-2026