FDA-Device2015-11-18Class II

Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy.

Boston Scientific Corporation
Hazard

Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully establishing telemetry during first scan of device when using a single-channel programmer. Telemetry can be established with additional scans; once established no further difficulty is expecte

Sold states
Distributed in the US to MA and PR.
Affected count
3
Manufactured in
4100 Hamline Ave N, N/A, Saint Paul, MN, United States
Products
Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy.