FDA-Device2023-11-29Class I
VariSoft infusion set, Model Number 1002828
Unomedical A/S
Hazard
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.
Sold states
US Nationwide distribution in the state of CA.
Affected count
1679 units
Manufactured in
Aaholmvej 1 - 3, OSTED, LEJRE, N/A, Denmark
Products
VariSoft infusion set, Model Number 1002828
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0233-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.