FDA-Device2025-10-29Class II
Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers
Baxter Healthcare Corporation
Hazard
Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.
Sold states
Distribution US nationwide, Canada and Colombia.
Affected count
5209
Manufactured in
1 Baxter Pkwy, Deerfield, IL, United States
Products
Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0233-2026Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief