FDA-Device2025-10-29Class II

Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers

Baxter Healthcare Corporation
Hazard

Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.

Sold states
Distribution US nationwide, Canada and Colombia.
Affected count
5209
Manufactured in
1 Baxter Pkwy, Deerfield, IL, United States
Products
Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0233-2026