FDA-Device2018-10-31Class II

Altrix¿ Precision Temperature Management System Model: 8001Product Usage: The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician or clinician with prescription privileges determines that temperature therapy is necessary or desirable.

Stryker Medical Division of Stryker Corporation
Hazard

Users may experience alarm fatigue due to frequent alarming relating to patient temperature deviation beyond 0.5¿C while in Automatic mode. 2. Users may experience a Remove from Use code 9 (RFU 9) fault condition relating to inadvertent flow alarms.3. Users may experience code RFU 27 fault condition relating to mechanical interference between the device fan and filter

Sold states
Worldwide - US Nationwide Distribution and the countries of Foreign: Australia, Canada, Netherlands, Switzerland and United Kingdom
Affected count
389 units
Manufactured in
3800 E Centre Ave, N/A, Portage, MI, United States
Products
Altrix¿ Precision Temperature Management System Model: 8001Product Usage: The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician or clinician with prescription privileges determines that temperature therapy is necessary or desirable.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0235-2019