FDA-Device2023-11-15Class II
VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)
Ortho-Clinical Diagnostics, INc.
Hazard
Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
Sold states
US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea
Affected count
342 units
Manufactured in
513 Technology Blvd, Rochester, NY, United States
Products
VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0236-2024Don't want to check this manually?