FDA-Device2025-10-29Class IIPROCESSING DEFECT

BD BACTEC 9120 blood culture system recalled for unauthorized credential access risk

OTHERNationwide distributionMedical devices

What this means

Unauthorized actors gained access to service credentials used by BD technical support teams for the BACTEC 9120 blood culture system. Until credentials are updated, there is a risk that unauthorized parties could access the system or associated patient data.

  • Check if your facility uses a BACTEC 9120 System (Catalog Numbers 445570, 445702, or 44557008)
  • Contact BD technical support for credential updates and security guidance
  • Follow any interim security measures provided by BD until updates are complete
The agency's original wording
Hazard

Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Intended for
All ages
Illnesses reported
None reported
Sold states
Worldwide distribution. US Nationwide including Guam and Puerto Rico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Burundi, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guatemala, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Mali, Malta, Mauritania, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Sierra Leone, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Timor-Leste, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zambia, and Zimbabwe.
Affected count
1,086 units
Manufactured in
7 Loveton Cir, Sparks, MD, United States
Products
BD BACTEC Blood Culture System, labeled as the following: 1. BACTEC 9120 System, Catalog Numbers: 445570, 445702. 2. INSTRUMENT BACTEC 9120 UNREPAIRED, Catalog Number: 44557008.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0237-2026