FDA-Device2012-11-14Class II
STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Diagnostica Stago, Inc.
Hazard
Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite¿.
Sold states
Worldwide Distribution - USA (nationwide) and Canada
Affected count
232 units US; 5 units Canada
Manufactured in
5 Century Dr, N/A, Parsippany, NJ, United States
Products
STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0252-2013Don't want to check this manually?