FDA-Device2020-10-28Class II

8 IN (20CM) APPX 1.3ml, EXT SET w/3-Port NanoClave Manifold, Check Valve.1 unit per pouch, 50 pouches per case. UDI:(01)10887709085814(17)250101(30)50(10)4605676. CLAVE Neutron is a normally closed, bidirectional connector intended for use as an accessory to an Intravascular catheter placed in the vein or artery. The device may be used for the administration of blood and fluids to patients, including pediatrics and immunocompromised patients. .

ICU Medical, Inc.
Hazard

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Sold states
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Affected count
Total of all products (Listed #1 thru 101) = 304735 units
Manufactured in
951 Calle Amanecer, N/A, San Clemente, CA, United States
Products
8 IN (20CM) APPX 1.3ml, EXT SET w/3-Port NanoClave Manifold, Check Valve.1 unit per pouch, 50 pouches per case. UDI:(01)10887709085814(17)250101(30)50(10)4605676. CLAVE Neutron is a normally closed, bidirectional connector intended for use as an accessory to an Intravascular catheter placed in the vein or artery. The device may be used for the administration of blood and fluids to patients, including pediatrics and immunocompromised patients. .

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0262-2021