FDA-Device2017-12-27Class II
The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF
Medacta Usa Inc
Hazard
Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.
Sold states
nationwide
Affected count
12,768 units
Manufactured in
3973 Delp St, Memphis, TN, United States
Products
The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0266-2018Don't want to check this manually?