FDA-Device2013-11-20Class II
Siemens syngo.plaza image processing system image processing radiological system
Siemens Medical Solutions USA, Inc
Hazard
The firm became aware of an unintended behavior when using the syngo.plaza. The "Rename" functionality can cause an unintended patient merge if the patient is sent to another system. This issue affects customers who transfer patient data between two or more syngo.plaza or syngo Imaging XS systems.
Sold states
Nationwide Distribution including MO, MA, NY, WI, OH, NY, CA, FL, NC, MN, and WA.
Affected count
19
Manufactured in
51 Valley Stream Pkwy, Malvern, PA, United States
Products
Siemens syngo.plaza image processing system image processing radiological system