FDA-Device2013-11-20Class II

Siemens syngo.plaza image processing system image processing radiological system

Siemens Medical Solutions USA, Inc
Hazard

The firm became aware of an unintended behavior when using the syngo.plaza. The "Rename" functionality can cause an unintended patient merge if the patient is sent to another system. This issue affects customers who transfer patient data between two or more syngo.plaza or syngo Imaging XS systems.

Sold states
Nationwide Distribution including MO, MA, NY, WI, OH, NY, CA, FL, NC, MN, and WA.
Affected count
19
Manufactured in
51 Valley Stream Pkwy, Malvern, PA, United States
Products
Siemens syngo.plaza image processing system image processing radiological system

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0271-2014

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief