FDA-Device2015-11-25Class II

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Precision Medical, Inc.
Hazard

These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).

Sold states
Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.
Affected count
315
Manufactured in
300 Held Dr, N/A, Northampton, PA, United States
Products
Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0272-2016