FDA-Device2021-12-01Class II

ARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF 01G06-11/1G06; ARCHITECT c16000 REF 03L77-01/3L77;

Abbott Laboratories
Hazard

Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes.

Sold states
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: AFGHANISTAN ALBANIA, ALGERIA, ANDORRA, ANGOLA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BELIZE, BENIN, BERMUDA, BOLIVIA, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, BURKINA FASO, BURUNDI, CAMBODIA, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEMOCRATIC REP. OF CONGO, DENMARK, FAROE ISLANDS, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ERITREA, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GEORGIA, GERMANY, GHANA, GREECE, GREENLAND, GUATEMALA, GUINEA, HAITI, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MADAGASCAR, MALAWI, MALAYSIA, MALDIVES, MALI, MALTA, MAURITANIA, MAURITIUS, MEXICO, MOLDOVA, MONTENEGRO, MONTSERRAT, MOROCCO, MOZAMBIQUE, MYANMAR, NAMIBIA, NEPAL, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, ST. LUCIA, ST. VINCENT, SURINAME, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TAJIKISTAN, TANZANIA, THAILAND, TRINIDAD & TOBAGO, TUNISIA, TURK & CAICOS, TURKEY, UAE, UGANDA, UK, UKRAINE, URUGUAY, UZBEKISTAN, VIETNAM, WEST BANK/GAZA STRIP, YEMEN, ZAMBIA, ZIMBABWE,
Affected count
7,535 systems
Manufactured in
1915 Hurd Dr, Irving, TX, United States
Products
ARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF 01G06-11/1G06; ARCHITECT c16000 REF 03L77-01/3L77;

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0272-2022