FDA-Device2018-01-03Class II

VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Ortho-Clinical Diagnostics
Hazard

The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.

Sold states
Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, Russia and Venezuela.
Affected count
US; 4680; Foreign: 1286
Manufactured in
513 Technology Blvd, N/A, Rochester, NY, United States
Products
VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0273-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief