FDA-Device2013-11-20Class II
Draeger Movita 603x/DVE 803x Ceiling Supply Unit. It is intended for use in operating theaters and intensive care
Draeger Medical, Inc.
Hazard
Cases were encountered with the Draeger Movita 603x/DVE 803x Ceiling Supply Units where the drive screw in the motor unit broke and the lift arm fell down approximately 60 cm (23.6 inches).
Sold states
Worldwide distribution: US (Nationwide)and in the country of Canada. (note: the ceiling supply units are not classified as a medical device by Health Canada).
Affected count
139
Manufactured in
3135 Quarry Rd, Telford, PA, United States
Products
Draeger Movita 603x/DVE 803x Ceiling Supply Unit. It is intended for use in operating theaters and intensive care
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0282-2014Don't want to check this manually?