FDA-Device2024-11-13Class II
Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Canon Medical System, USA, INC.
Hazard
When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.
Sold states
US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA
Affected count
27
Manufactured in
2441 Michelle Dr, N/A, Tustin, CA, United States
Products
Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0284-2025Don't want to check this manually?