FDA-Device2015-11-25Class II
Prism Medical P-300 Portable Consumer Lift; PN 303050, 303050R, and 303051; weight specification 300 lbs. /136 kg.
ErgoSafe Products, LLC (DBA) Prism Medical
Hazard
Drive shaft and over speed arm may simultaneously fail potentially resulting in a patient falling during use
Sold states
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, South Korea, Taiwan, Mexico, Japan, and Great Britain.
Affected count
1,721 lifts
Manufactured in
10888 Metro Ct, N/A, Maryland Heights, MO, United States
Products
Prism Medical P-300 Portable Consumer Lift; PN 303050, 303050R, and 303051; weight specification 300 lbs. /136 kg.