FDA-Device2018-11-07Class II

Magellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood. Catalog Number: 70-6762 Primary DI No: 00850355006000 The LeadCare II Blood Lead Test Kit is for in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and treatment reagent of the test kit are specific for lead only. Contents of the LeadCare II Blood Lead Test Kit must only be us

Magellan Diagnostics, Inc.
Hazard

Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 ¿g/dL). Results for patient testing with this lot will be displayed as HIGH.

Sold states
Worldwide distribution: US (Nationwide) and countries of: Spain and South Africa.
Affected count
1,465 kits
Manufactured in
101 Billerica Ave Bldg 4, N/A, North Billerica, MA, United States
Products
Magellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood. Catalog Number: 70-6762 Primary DI No: 00850355006000 The LeadCare II Blood Lead Test Kit is for in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and treatment reagent of the test kit are specific for lead only. Contents of the LeadCare II Blood Lead Test Kit must only be used with a LeadCare II Analyzer.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0285-2019