FDA-Device2012-11-21Class II
Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.
Stryker Howmedica Osteonics Corp.
Hazard
Stryker has become aware that insufficient taper torsional strength may result when a PAR extension piece is used with a forged CoCr stem.
Sold states
Worldwide Distribution - US Nationwide
Affected count
301 units
Manufactured in
325 Corporate Drive, Mahwah, NJ, United States
Products
Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0286-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief