FDA-Device2012-11-21Class II

Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.

Stryker Howmedica Osteonics Corp.
Hazard

Stryker has become aware that insufficient taper torsional strength may result when a PAR extension piece is used with a forged CoCr stem.

Sold states
Worldwide Distribution - US Nationwide
Affected count
301 units
Manufactured in
325 Corporate Drive, Mahwah, NJ, United States
Products
Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0286-2013