FDA-Device2024-11-13Class II

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515

Siemens Healthcare Diagnostics Inc
Hazard

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.

Sold states
US Nationwide. Canada.
Affected count
1292 units
Manufactured in
2 Edgewater Dr, N/A, Norwood, MA, United States
Products
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0286-2025