FDA-Device2015-11-25Class II
Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.
Fresenius Medical Care Renal Therapies Group, LLC
Hazard
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Sold states
Nationwide Distribution.
Affected count
348 units
Manufactured in
920 Winter St, N/A, Waltham, MA, United States
Products
Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.
Official notice