FDA-Device2012-11-21Class II
ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.
Zimmer, Inc.
Hazard
Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the
Sold states
Worldwide Distribution - USA (nationwide) and Internationally to Belgium, Czech Republic, Egypt, France, Germany, Morocco, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey, and UAE.
Affected count
204
Manufactured in
345 E Main St, Warsaw, IN, United States
Products
ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0296-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief