FDA-Device2018-11-07Class II
Ysio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.
Siemens Medical Solutions USA, Inc
Hazard
In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.
Sold states
The products were distributed to the following US states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, and WY.
Affected count
142
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Ysio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.