FDA-Device2018-11-07Class II
IQon Spectral CT (Model No. 728332), Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.
Philips Medical Systems (Cleveland) Inc
Hazard
It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple analyses of the amount of plaque detected, as if they all apply to the patient at the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.
Sold states
US Nationwide Distribution
Affected count
64
Manufactured in
595 Miner Rd, N/A, Cleveland, OH, United States
Products
IQon Spectral CT (Model No. 728332), Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0306-2019Don't want to check this manually?