FDA-Device2024-11-20Class I
microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide a microelectrode or instrument in the brain during stereotactic functional neurosurgical procedures.
FHC, Inc.
Hazard
A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
Sold states
Domestic distribution only to the following states: AZ, CA, PA, TN, TX, WA.
Affected count
8 packs of 5ea (40 total)
Manufactured in
1201 Main St, N/A, Bowdoin, ME, United States
Products
microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide a microelectrode or instrument in the brain during stereotactic functional neurosurgical procedures.