FDA-Device2016-11-02Class II
Zimmer¿ Gender Solutions" Patellofemoral Joint Prosthesis Milling Handpiece
Zimmer Biomet, Inc.
Hazard
Complaints have been reported that the handpiece is inoperable. The investigation found that the complaints were related to the lack of preventative maintenance contrary to what is recommended in the instructions for use included with the handpiece. An inoperable device may lead to a delay in surgery
Sold states
Nationwide. CANADA HONG KONG KOREA SINGAPORE CHINA VENEZUELA AUSTRALIA JAPAN NETHERLANDS GERMANY INDIA TAIWAN
Affected count
703
Manufactured in
1800 W Center St, Warsaw, IN, United States
Products
Zimmer¿ Gender Solutions" Patellofemoral Joint Prosthesis Milling Handpiece
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0317-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief