FDA-Device2020-11-11Class II
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm Item Number: 00-8065-540-22 Hip prosthesis component
Zimmer Biomet, Inc.
Hazard
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
Sold states
US Nationwide
Affected count
412 units total
Manufactured in
1800 W Center St, N/A, Warsaw, IN, United States
Products
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm Item Number: 00-8065-540-22 Hip prosthesis component
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0323-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief