FDA-Device2021-12-15Class II
Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA
Covidien Llc
Hazard
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
Sold states
Brazil
Affected count
2 units OUS
Manufactured in
15 Hampshire St, N/A, Mansfield, MA, United States
Products
Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0332-2022Don't want to check this manually?