FDA-Device2021-12-15Class II

Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA

Covidien Llc
Hazard

Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit

Sold states
Brazil
Affected count
2 units OUS
Manufactured in
15 Hampshire St, N/A, Mansfield, MA, United States
Products
Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0332-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief