FDA-Device2023-11-29Class II
Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-VT3.
Aizu Olympus Co., Ltd.
Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Sold states
Worldwide - US Nationwide distribution.
Affected count
46 units
Manufactured in
3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, Japan
Products
Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-VT3.
Official notice