FDA-Device2023-11-29Class II
Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.
Aizu Olympus Co., Ltd.
Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Sold states
Worldwide - US Nationwide distribution.
Affected count
3627 units
Manufactured in
3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, Japan
Products
Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0334-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.