FDA-Device2015-12-02Class II

NxStage System One S Cycler - Model no. NX1000-3. For home hemodialysis.

NxStage Medical, Inc.
Hazard

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Sold states
Worldwide Distribution.
Affected count
2,134 units
Manufactured in
350 Merrimack Street, N/A, Lawrence, MA, United States
Products
NxStage System One S Cycler - Model no. NX1000-3. For home hemodialysis.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0335-2016

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief