FDA-Device2023-11-29Class II

Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.

Aizu Olympus Co., Ltd.
Hazard

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Sold states
Worldwide - US Nationwide distribution.
Affected count
27 units
Manufactured in
3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, Japan
Products
Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0336-2024