FDA-Device2023-11-29Class II
Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.
Aizu Olympus Co., Ltd.
Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Sold states
Worldwide - US Nationwide distribution.
Affected count
0 units
Manufactured in
3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, Japan
Products
Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.