GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.
The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.