FDA-Device2021-12-15Class I

Puritan Bennett 980 Series Ventilator, 980X1ENNISB

Covidien Llc
Hazard

Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

Sold states
Worldwide distribution - US Nationwide.
Affected count
278 total
Manufactured in
15 Hampshire St, N/A, Mansfield, MA, United States
Products
Puritan Bennett 980 Series Ventilator, 980X1ENNISB

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0348-2022