FDA-Device2023-11-29Class II

GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007

Instrumentation Laboratory
Hazard

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Sold states
US States: MA, MN, NY. India.
Affected count
2448 units
Manufactured in
180 Hartwell Rd, N/A, Bedford, MA, United States
Products
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0358-2024