FDA-Device2018-11-07Class II

FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) (same product) FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre- Analytical and Post-Analytical processing,

Inpeco S.A.
Hazard

The firm has identified customer sites where there are modules on which the safety pacemaker label is missing. The information regarding the label included in the Operations Manual has been updated to specify that: 1) the risk of a pacemaker malfunction applies to anyone with a pacemaker who may work on these modules; and 2) the minimum safety distance has been increased to 200 mm (7.87") from the transport mechanisms. Any person with a pacemaker must not get closer than this distance from the modules axes.

Sold states
Foreign distribution to Netherlands, Italy, Belgium, Austria, France, Spain, and Germany.
Affected count
679
Manufactured in
Via San Gottardo 10, N/A, Lugano, N/A, Switzerland
Products
FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) (same product) FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre- Analytical and Post-Analytical processing,

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0361-2019