FDA-Device2023-11-29Class II
Ventlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3. RescuMed Pediatric Resuscitator REF #s: BVMB812, BVMB812-F, BVMB810, BVMB810-F, BVMB810S-F.
SunMed Holdings, LLC
Hazard
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Sold states
US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
Affected count
N/A
Manufactured in
2710 Northridge Dr Nw Ste A, N/A, Grand Rapids, MI, United States
Products
Ventlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3. RescuMed Pediatric Resuscitator REF #s: BVMB812, BVMB812-F, BVMB810, BVMB810-F, BVMB810S-F.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0364-2024Don't want to check this manually?