FDA-Device2016-11-16Class II

The surgical table is a stand alone unit. Three (3) floor locking cylinders are attached to the base of the surgical table. The individual cylinders are replaceable. Product Usage: The cylinders are used to lock the surgical table in place.

Schaerer Medical USA Inc
Hazard

The floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect "O" ring which may cause the cylinder to leak.

Sold states
Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed
Affected count
4 surgical tables with defective floor locking cylinders were distributed. Each table contains 3 cylinders for a total of 12 cylinders.
Manufactured in
675 Wilmer Ave, Cincinnati, OH, United States
Products
The surgical table is a stand alone unit. Three (3) floor locking cylinders are attached to the base of the surgical table. The individual cylinders are replaceable. Product Usage: The cylinders are used to lock the surgical table in place.