FDA-Device2018-01-24Class II
RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
RAYSEARCH LABORATORIES AB
Hazard
For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.
Sold states
US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
Affected count
117 units
Manufactured in
Sveavaegen 9, N/A, Stockholm, N/A, Sweden
Products
RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0380-2018Don't want to check this manually?