FDA-Device2013-12-04Class II

REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device, Single Use Only. Designed to release the transverse carpal ligament.

Biomet, Inc.
Hazard

Security Blades were manufactured oversized and potentially won't advance or may be difficult to advance through the mating Security Clip.

Sold states
Worldwide Distribution: USA (nationwide) and Internationally to Japan.
Affected count
1130
Manufactured in
56 E Bell Dr, N/A, Warsaw, IN, United States
Products
REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device, Single Use Only. Designed to release the transverse carpal ligament.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0381-2014