FDA-Device2019-11-20Class II
ORTHOFIX Catalogue Number: ref 1-1100201,DRILL BIT Diameter .4,8 MM L.240 MM TIN COATED - QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18033509856068
Orthofix Srl
Hazard
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sold states
US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
Affected count
N/A
Manufactured in
Via Delle Nazioni 9 O 73, N/A, Veronaverona, N/A, Italy
Products
ORTHOFIX Catalogue Number: ref 1-1100201,DRILL BIT Diameter .4,8 MM L.240 MM TIN COATED - QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18033509856068
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0381-2020Don't want to check this manually?